| 1 |
- Prepare canal using preferred
method.
- Choose a Quick-Fill™ two sizes
smaller than the last file used to prepare apex.
- Set endostop on the Quick-Fill™ to
1 mm short of the apex. (the mm markings on the Quick-Fill™ corresponds to 18mm, 19mm, 20mm and 24mm.)
|
2 | Lightly coat or dip Quick-Fill™ in sealer of choice. Before starting the handpiece, position the Quick-Fill™ at the orifice until a light resistance is felt.DO NOT apply any pressure or force. |
| 3 |
- Start handpiece, in clockwise rotation
between 3000-6000 rpm, and apply light pressure. The gutta-percha will give a tactile sense of plasticizing.
- The Quick -Fill TM MUST be used in
CLOCKWISE rotation ONLY.
|
4 |
- Designed as an inverted ‘K’ type file, this titanium carrier forces the plasticized gutta-percha downward as it is withdraw from canal.
|
| 5 |
- Proceed apically using constant light
pressure until working length is reached (use endostop).
- Once working lenght is reached,
continue rotation and slowly remove the carrier from canal.
- DO NOT remove carrier too fast as
it may leave a void.
|
6 |
- Last, take X-Ray and verify fill to the apex.
- If apex is not sealed, adjust endostop (longer lenght) and repeat procedure with the same Quick-FillTM instrument
|