QUICK-FILL Instructions

1
  • Prepare canal using preferred
    method.
  • Choose a Quick-Fill™ two sizes
    smaller than the last file used to
    prepare apex.
  • Set endostop on the Quick-Fill™ to
    1 mm short of the apex. (the
    mm markings on the Quick-Fill™
    corresponds to 18mm, 19mm,
    20mm and 24mm.)
2Lightly coat or dip Quick-Fill™ in
sealer of choice.
Before starting the handpiece,
position the Quick-Fill™ at the orifice
until a light resistance is felt.

DO NOT apply any pressure or force.

3
  • Start handpiece, in clockwise rotation
    between 3000-6000 rpm, and apply light
    pressure. The gutta-percha will give a
    tactile sense of plasticizing.
  • The Quick -Fill TM MUST be used in
    CLOCKWISE rotation ONLY.
4
  • Designed as an inverted ‘K’ type file, this titanium carrier forces the plasticized gutta-percha downward as it is withdraw from canal.
5
  • Proceed apically using constant light
    pressure until working length is
    reached (use endostop).
  • Once working lenght is reached,
    continue rotation and slowly remove
    the carrier from canal.
  • DO NOT remove carrier too fast as
    it may leave a void.
6
  • Last, take X-Ray and verify fill to the apex.
  • If apex is not sealed, adjust endostop (longer lenght) and repeat procedure with the same Quick-FillTM instrument